Cooling water
Filters
Last update 28 Jan 2020
- Rope filter 25 um x4
- Rope filter 10 um x5
NEW Water Cooling System
At the moment our water system has three outputs: A - B - C.
The incoming water in filtered, before the outputs, at 25 um.
In addition the outputs have the following dedicated filters:
A-> 10 um || B -> 20 um || C -> 10 um
Device | Temperature (°C) | Flow (l/min) | Pressure (bar) | Max Output Temp. | Power dissipation1 | Water use | Internal element | Max filter size |
Output A: actual max flux 4 l/min |
InnoLas SL 1200 | 20 (MAX) | 8.0 (MIN) | 2 - 6 | 28 | 3,0 kW | cooling internal closed circuits | heat exchanger+2mm hole | 50 um |
Output B: actual max flux 8 l/min |
Quanta-Ray pulsed Nd:YAG PRO 230 | 15-23 | 7.6 (@ 30/50Hz or 3.8 @ 10Hz) | 2.8 - 4.1 (40psi - 60psi) | 28 | 3,5 kW | cooling internal closed circuits | heat exchanger | 20um |
Output C: actual flux 2l/min |
Dye circulator (Radiant Dye) | 15-25 | 1 | 2 | x 22 | 200 W |
TurboPumps Pfeiffer 300 HiPace | 15-35 | 0.8 | 2 | x 22 | 60 W |
TurboPumps Pfeiffer 300 HiPace | 15-35 | 0.8 | 2 | x 22 | 60 W |
TurboPumps Pfeiffer 300 HiPace | 15-35 | 0.8 | 2 | x 22 | 60 W |
TurboPumps Pfeiffer 521P (YP) | 15-25 | 1.7 | 2 | x 22 | 60 W |
TurboPumps Pfeiffer 521P (YP) | 15-25 | 1.7 | 2 | x 22 | 60 W |
TurboPumps Pfeiffer 521P (YP) | 15-25 | 1.7 | 2 | x 22 | 60 W |
TurboPumps Pfeiffer TPU 240 | ~15-25 | ? | ~2 | x 22 | 60 W |
At the moment not connected to any of the outputs |
ThermoTek Chiller P315 | 5-25 | 5-10 | ? | ? | ~1,5 kW | cooling internal closed circuits | heat exchanger | 50 um |
1: The power is calculated assuming incoming water at 17°C
x:These devices are connected in series. Output temperature is due to the power dissipation of all of them.
Real Power consumption:
Output | Flux | Pressure | Power to dissipate |
A | 4 l/min | 3 bar | 3.0 kW |
B | 8 l/min | 4bar | 3.5 kW |
C | 2 l/min | 2bar | 620W |
Other | 5 l/min | 4 bar | ~1,5kW |
Total incoming flow: 24 l/min
Incoming pressure: 6bar
Total real power to dissipate via water: ~ 8,7 kW\\